Malfunctions / Troubleshooting
Fault finding list
Switchpoint "Pump OFF" exceeded - no function
Faul t: Mains voltage is not applied.
Remedy:Switch on mains voltage. Connect level switch according to wiring diagram
(Fig. 5).
Faul t: The electrical conductivity is too low.
Remedy:Set code switch to 0.5 µS/cm.
Faul t: The earth connection to the vessel is interrupted.
Remedy:Clean seating surfaces (see Fig. 3) and insert metal joint ring Ø33x39
DIN 7603 (made from s. s. 1.4301). Do not insulate the level switch with
hemp or PTFE tape!
Faul t: Electronic circuit board defective.
Remedy:Replace electronic circuit board.
Level below switchpoint "Low level" - no function
Faul t: The electrode tips have earth contact.
Remedy:Check and, if necessary, change position of installation.
Faul t: If the equipment is installed inside the boiler: The vent hole in the protection
tube does not exist or is obstructed.
Remedy:Check installation of level switch. Make sure that the level in the protection
tube corresponds to the actual water level (see Examples of Installation).
Switchpoint has been reached - incorrect function
Faul t: The switching function has not been assigned correctly. The electrode tips
have been cut to the wrong length.
Remedy:Assign the electrode tips correctly and change connections on the electronic
circuit board correspondingly.
Faul t: The switching function does not correspond to the selected function. The
code switch has been set to the wrong value.
Remedy:Set code switch according to the selected function.