MX/QX HART Field Unit FCD LMENIM2340-00 – 1/14
flowserve.com
Command Response Code
No. Description Code Class Description
63 Read Analog Channel
Information
0
1
2
3-4
5
6
7-15
16
17-127
Success
N/A
Error
N/A
Error
Error
N/A
Error
N/A
No Command-Specific Errors
Undefined
Invalid Selection
Undefined
Too Few Data Bytes Received
Device-Specific Command Error
Undefined
Access Restricted
Undefined
65 Write Analog Channel Range
Values
0
1-4
5
6
7
8
9
10
11
12
13
14
15
16
17-27
28
29
17-31
32
33-127
Success
N/A
Error
Error
Error
Warning
Error
Error
Error
Error
Error
Warning
Error
Error
N/A
Error
Error
N/A
Error
N/A
No Command-Specific Errors
Undefined
Too Few Data Bytes Received
Device-Specific Command Error
In Write Protect Mode
Set to Nearest Possible Value (Upper or Lower Range Pushed)
Lower Range Value Too High
Lower Range Value Too Low
Upper Range Value Too High
Upper Range Value Too Low
Upper and Lower Range Values Out of Limits
Span Too Small (Device Accuracy May be Impaired)
Invalid Analog Channel Code Number
Access Restricted
Undefined
Invalid Range Units Code
Invalid Span
Undefined
Busy
Undefined
66 Enter/Exit Fixed Analog Channel
Mode
0
1-2
3
4
5
6
7
8-10
11
12
13-14
15
16
17-31
32
33-127
Success
N/A
Error
Error
Error
Error
Error
N/A
Error
Error
N/A
Error
Error
N/A
Error
N/A
No Command-Specific Errors
Undefined
Passed Parameter Too Large
Passed Parameter Too Small
Too Few Data Bytes Received
Device-Specific Command Error
In Write Protect Mode
Undefined
In Multi-drop Mode
Invalid Units Code
Undefined
Invalid Analog Channel Code Number
Access Restricted
Undefined
Busy
Undefined
67 Trim Analog Channel Zero 0
1-2
3
4
5
6
7
8
9
10
11
12
13-14
15
16
17-31
32
33-127
Success
N/A
Error
Error
Error
Error
Error
N/A
Error
N/A
Error
Error
N/A
Error
Error
N/A
Error
N/A
No Command-Specific Errors
Undefined
Passed Parameter Too Large
Passed Parameter Too Small
Too Few Data Bytes Received
Device-Specific Command Error
In Write Protect Mode
Undefined
Not in Proper Analog Channel Mode
Undefined
In Multi-drop Mode
Invalid Units Cod
Undefined
Invalid Analog Channel Code Number
Access Restricted
Undefined
Busy
Undefined
68 Trim Analog Channel Gain 0
1-2
3
4
5
6
7
8
9
10
11
12
13-14
15
16
17-31
32
33-127
Success
N/A
Error
Error
Error
Error
Error
N/A
Error
N/A
Error
Error
N/A
Error
Error
N/A
Error
N/A
No Command-Specific Errors
Undefined
Passed Parameter Too Large
Passed Parameter Too Small
Too Few Data Bytes Received
Device-Specific Command Error
In Write Protect Mode
Undefined
Not in Proper Analog Channel Mode
Undefined
In Multi-drop Mode
Invalid Units Cod
Undefined
Invalid Analog Channel Code Number
Access Restricted
Undefined
Busy
Undefined